An LC-MS chromatogram plots a detector response over time. A mass spectrum plots ion abundance against m/z. They are different views of analytical data. A peak at 8.2 minutes and a peak at m/z 651.3 describe different coordinates, even if both appear on the same report page.
This guide is for reading plots and preparing a clearer document request. It does not calculate peptide purity, assess a supplied batch or replace a laboratory's interpretation. Read the axes and trace label before comparing shapes.
Two dimensions, plus the signal intensity
Shimadzu's LC-MS reference describes retention time, m/z and ion intensity as the dataset's three dimensions. Taking a view along time produces a chromatographic trace; a mass-spectrum view concerns the m/z distribution for selected data. Read the manufacturer data-view explanation.
| View | Horizontal axis | What a plotted peak locates |
|---|---|---|
| Chromatogram | Time, commonly minutes | A response within the separation run |
| Mass spectrum | m/z | An ion signal at a mass-to-charge coordinate |
| Selected spectrum from a chromatographic region | m/z, with its time selection stated | Ion signals for that selected data region |
A useful screenshot retains both the axis labels and the selection context. Cropping away the time range or scan label can remove the information needed to understand which spectrum belongs with a chromatographic feature.
TIC, base-peak and extracted-ion traces
Thermo Fisher's software glossary distinguishes a total-ion-current chromatogram, which sums detected ion signals for each scan, from a base-peak chromatogram, which follows the most intense signal in each spectrum. Both are time traces, despite summarising the underlying spectra differently. Read the primary software glossary.
An extracted-ion chromatogram follows a selected m/z value or defined window over time. The selection belongs with the trace. A narrow extracted view and a total-ion view are not identical measurements merely because both horizontal axes show minutes. See the extracted-ion description.
A fictional report-reading example
Imagine a time-trace feature centred near 8.2 minutes. A separate panel shows a spectrum selected from that region, with a labelled signal at m/z 651.3. These numbers are invented for this explanation and are not an Aurelis test result.
A precise note would say: “The selected 8.2-minute region contains a labelled m/z 651.3 signal.” Saying “the peptide has a mass of 651.3” would add assumptions about charge, ion type and identification that those two labels alone do not establish. The charge-state guide develops that next distinction.
Why one display is not a complete identity result
A spectrum can contain several signals, while a time-trace feature can be viewed using different selections. Keep the laboratory's assignment separate from the plot location. Record whether a label identifies an observed ion, a proposed molecular species or a software annotation.
The largest displayed peak should not automatically become a percentage of peptide content. Here, the task is to identify the view being shown. Quantitative conclusions require their own defined method, response basis and interpretation; they are outside this plot-reading example.
Australian laboratory context without assumed affiliation
The University of Queensland's chemistry mass-spectrometry facility describes LC-MS/MS as providing separation and identification, alongside distinct accurate-mass, fragmentation and quantification capabilities. Its own description shows why an analytical request needs a defined question. It is a public technical reference, not an Aurelis testing-partner claim. Read the UQ facility description.
For an Australian documentation brief, describe the result you need to understand rather than requesting an unlabeled “LC-MS picture”. Ask for the trace type, axes, selected time interval, mass-spectral view and the laboratory's stated assignment in the same record.
Prepare a useful analytical-document brief
Retain the sample or report identifier, analysis date, method reference and original plot labels. Identify whether the spectrum is a single scan or a stated selection across time, and whether processing or background treatment is described. If a field is absent, list it as a question rather than filling it in yourself.
For a general document or laboratory-presentation requirement, use the laboratory specification enquiry. Describe the document question and Australian destination where relevant. That enquiry asks whether the team can assist; it does not establish that a test has been performed or a request accepted.